TítuloCentral corneal thickness measured with three optical devices and ultrasound pachometry
Autor(es)González-Pérez, Javier
González-Méijome, José Manuel
Rodríguez Ares, María T.
Parafita, Manuel A.
Palavras-chavePachometry
Central corneal thickness
Pentacam
Ultrasound
Orbscan
Specular microscopy
Bland
Altman
Bland-Altman
Data2011
EditoraLippincott, Williams & Wilkins
RevistaEye & Contact Lens: Science and Clinical Practice
Resumo(s)Purpose: To evaluate the reliability of three noninvasive pachometry methods against the ultrasound pachometer considered the gold standard. Methods: Central corneal thickness (CCT) was measured using a Paxis ultrasound (US) pachometer, Orbscan II, Pentacam, and Topcon SP-3000 specular microscope in 22 right eyes of 22 healthy adults (7 men, 15 women). Three repeated measures were obtained and then compared to obtain the repeatability of each instrument and the agreement between pachometers. Results: Pentacam is the optical system that renders values of CCT closer to those obtained with US pachometry. Conversely, a specular microscope shows a poorer agreement with US measures, and differences depend on the thickness being measured. The Orbscan system shows a consistent trend toward underestimation of CCT compared with US and Pentacam irrespective of the value measured. Conclusion: Different optical methods used to measure CCT showed significantly different results compared with US pachometry except for the Pentacam system. Clinicians should take into account the fact that specular microscopy might either underestimate or overestimate values of CCT, whereas Orbscan systematically overestimates CCT compared with US and Pentacam.
TipoArtigo
URIhttps://hdl.handle.net/1822/17172
DOI10.1097/ICL.0b013e31820c6ffc
ISSN1542-2321
1542-233X
Versão da editorahttp://journals.lww.com/
Arbitragem científicayes
AcessoAcesso restrito UMinho
Aparece nas coleções:ICVS - Artigos em revistas internacionais / Papers in international journals
CDF - OCV - Artigos/Papers (with refereeing)

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
5.5.58. GONZALEZ-PEREZ_11_E&L_ (Comparison of Pachometers).pdf
Acesso restrito!
277,46 kBAdobe PDFVer/Abrir

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID