Título: | Cost-efective lab-on-a-chip device for seawater pH quantification by optical methods |
Autor(es): | Pinto, Vânia Cristina Gonçalves Sousa, Paulo Jorge Teixeira Magalhaes, Vitor H. Araújo, Catarina Minas, Graça Gonçalves, L. M. |
Palavras-chave: | meta-Cresol Purple Microfluidic pH sensor seawater spectrophotometry |
Data: | 2019 |
Editora: | Institute of Electrical and Electronics Engineers Inc. |
Resumo(s): | This paper presents a novel lab-on-a-chip (LOC) device for seawater pH quantification based on spectrophotometric methods. The device integrates optical, electronics and microfluidic systems, providing an autonomous, compact and portable platform for in-situ, long-term and low-cost seawater pH measurements. The developed system was validated in laboratory with measurement of various seawater samples, of different pH, within the seawater pH interval (7.50-8.30). The response of the LOC shows a high reproducibility and stability with a resolution of 0.01 pH units. Thus, the device has a high potential to be use in moored oceanic applications. |
Tipo: | Artigo em ata de conferência |
URI: | https://hdl.handle.net/1822/68299 |
ISBN: | 9781728120072 |
DOI: | 10.1109/TRANSDUCERS.2019.8808810 |
Versão da editora: | https://ieeexplore.ieee.org/document/8808810 |
Arbitragem científica: | yes |
Acesso: | Acesso restrito UMinho |
Aparece nas coleções: | CMEMS - Artigos em livros de atas/Papers in proceedings |