TítuloApplication of a lab-made electronic nose for extra virgin olive oils commercial classification according to the perceived fruitiness intensity
Autor(es)Teixeira, Guilherme G.
Dias, Luís G.
Rodrigues, Nuno
Marx, Ítala M. G.
Veloso, Ana C. A.
Pereira, José A.
Peres, António M.
Palavras-chaveOlive oil
Fruity intensity commercial grade
Electronic nose
Metal oxide sensors
Linear discriminant analysis
Simulated annealing algorithm
DataMai-2021
EditoraElsevier
RevistaTalanta
CitaçãoTeixeira, Guilherme G.; Dias, Luís G.; Rodrigues, Nuno; Marx, Ítala M. G.; Veloso, Ana C. A.; Pereira, José A.; Peres, António M., Application of a lab-made electronic nose for extra virgin olive oils commercial classification according to the perceived fruitiness intensity. Talanta, 226(122122), 2021
Resumo(s)An electronic nose, comprising nine metal oxide sensors, has been built aiming to classify olive oils according to the fruity intensity commercial grade (ripely fruity or light, medium and intense greenly fruity), following European regulated complementary terminology. The lab-made sensor device was capable to differentiate standard aqueous solutions (acetic acid, cis-3-hexenyl, cis-3-hexen-1-ol, hexanal, 1-hexenol and nonanal) that mimicked positive sensations (e.g., fatty, floral, fruit, grass, green and green leaves attributes) and negative attributes (e.g., sour and vinegary defects), as well as to semi-quantitatively classify them according to the concentration ranges (0.05 to 2.25 mg/kg). For that, unsupervised (principal component analysis) and supervised (linear discriminant analysis: sensitivity of 92% for leave-one-out cross validation) classification multivariate models were established based on nine or six gas sensors, respectively. It was also showed that the built E-nose allowed differentiating/discriminating (sensitivity of 81% for leave-one-out cross validation) extra virgin olive oils according to the perceived intensity of fruitiness as ripely fruity, light, medium or intense greenly fruity. In conclusion, the gas sensor device could be used as a practical preliminary non-destructive tool for guaranteeing the correctness of olive oil fruitiness intensity labelling.
TipoArtigo
URIhttps://hdl.handle.net/1822/70102
DOI10.1016/j.talanta.2021.122122
ISSN0039-9140
Versão da editorahttps://www.sciencedirect.com/science/article/pii/S0039914021000436#!
Arbitragem científicayes
AcessoAcesso aberto
Aparece nas coleções:CEB - Publicações em Revistas/Séries Internacionais / Publications in International Journals/Series

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
document_54125_1.pdf3,9 MBAdobe PDFVer/Abrir

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID