dc.description.abstractThis paper describes a 64 pixels X-ray imager architecture and fabrication process. The imager is composed by a 8×8 photodiodes array, fabricated in CMOS process, and an array of wells £lled with scintillating crystals. A thick-£lm of aluminum is etched in order to achieve square wells with 500 um depth. The wells are filled with CsI:Tl scintillating crystals and placed above the photodiodes. The scintillating crystals convert the X-ray energy into visible light, which is guided into the photodiodes by the re¤ective aluminum walls, avoiding crosstalk between adjacent detectors. Usually, the spatial resolution of the scintillating x-rays detectors is identical to the scintillators thickness. By using the light guides, the scintillator thickness can be adjusted in order to achieve optimal absorption ef£ciency, since the spatial resolution is established by the pixel size (200 um side).eng
dc.language.isoengeng
dc.publisherUniversidade do Minhoeng
dc.rightsopenAccesseng
dc.subjectX-rayeng
dc.subjectDigital radiologyeng
dc.subjectScintillatoreng
dc.subjectEtchingeng
dc.titleCMOS X-ray imager for dental radiographyeng
dc.typeconferencePapereng
dc.peerreviewedyeseng
Aparece nas coleções:DEI - Artigos em atas de congressos internacionais

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
MP40.pdf1,06 MBAdobe PDFVer/Abrir

Partilhe no FacebookPartilhe no TwitterPartilhe no DeliciousPartilhe no LinkedInPartilhe no DiggAdicionar ao Google BookmarksPartilhe no MySpacePartilhe no Orkut
Exporte no formato BibTex mendeley Exporte no formato Endnote Adicione ao seu ORCID