Design of a pressure sensor for monitoring of post-endovascular aneurysm repair
| dc.contributor.author | Sepúlveda, A. T. | |
| dc.contributor.author | Santos, Isa C. T. | |
| dc.contributor.author | Fachin, F. | |
| dc.contributor.author | Guzmán de Villoria, R. | |
| dc.contributor.author | Wardle, B. L. | |
| dc.contributor.author | Pontes, A. J. | |
| dc.contributor.author | Viana, J. C. | |
| dc.contributor.author | Rocha, Luís A. | |
| dc.date.accessioned | 2011-11-29T16:49:53Z | |
| dc.date.available | 2011-11-29T16:49:53Z | |
| dc.date.issued | 2011-01-26 | |
| dc.description.abstract | This paper introduces the design and fabrication process of a flexible pressure sensor for monitoring postendovascular aneurysm repairs (EVAR). Biocompatible flexible PDMS membranes with embedded aligned carbon nanotubes (CNTs) with a conductivity of 11 S.m-1 and elastic modulus of 2 MPa are used to build a LC network for passive telemetry readout-out. The pressure sensor has a modelled sensitivity 14fF/mmHg for a pressure range between 6-47 mmHg, in agreement with the required application. The pressure sensor, with a 250 μm thickness and total area of 1 cm2, will be contained within the stent-graft and used to measure the pressure inside the aneurysm sac to detect complications related to the EVAR procedure. | por |
| dc.description.sponsorship | Fundação para a Ciência e a Tecnologia (FCT) - SFRH/BD/42922/2008, MIT-Pt/EDAM-EMD/0007/2008 | por |
| dc.distribution | international | |
| dc.identifier.isbn | 9789898425379 | por |
| dc.identifier.uri | https://hdl.handle.net/1822/14646 | |
| dc.language.iso | eng | por |
| dc.peerreviewed | yes | por |
| dc.publisher | SCITEPRESS | por |
| dc.rights | restrictedAccess | por |
| dc.subject | Implantable pressure sensor | por |
| dc.subject | Aneurysms | por |
| dc.subject | Carbon nanotubes | por |
| dc.subject | Pressure sensor design | por |
| dc.subject.wos | Science & Technology | por |
| dc.title | Design of a pressure sensor for monitoring of post-endovascular aneurysm repair | por |
| dc.type | conferencePaper | |
| dspace.entity.type | Publication | en |
| oaire.citationConferenceDate | 26 - 29 Janeiro 2011 | por |
| oaire.citationConferencePlace | Roma, Itália | por |
| oaire.citationEndPage | 22 | por |
| oaire.citationStartPage | 14 | por |
| oaire.citationTitle | International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2011) | por |
| sdum.bookTitle | BIODEVICES 2011 | por |
| sdum.conferencePublication | International Joint Conference on Biomedical Engineering Systems and Technologies (BIOSTEC 2011) | por |
| sdum.publicationstatus | published | por |