Ti6Al4V laser surface preparation and functionalization using hydroxyapatite for biomedical applications

dc.contributor.authorFaria, D.por
dc.contributor.authorAbreu, C. S.por
dc.contributor.authorBuciumeanu, Mihaelapor
dc.contributor.authorDourado, N.por
dc.contributor.authorCarvalho, Óscar Samuel Novaispor
dc.contributor.authorSilva, F. S.por
dc.contributor.authorMiranda, G.por
dc.date.accessioned2018-03-12T15:10:06Z
dc.date.issued2018
dc.description.abstractThis work presents a novel texture design for implants surface functionalization, through the creation of line-shaped textures on Ti6Al4V surfaces and subsequent sintering of hydroxyapatite (HAp) powder into the designated locations. HAp-rich locations were designed to avoid HAp detachment during insertion, thus guaranteeing an effective osseointegration. This process starts by creating textured lines using a Nd:YAG laser, filling these lines with HAp powder and sintering HAp using a CO2laser. The adhesion of HAp is known to be influenced by HAp sintering parameters, especially laser power and scanning speed and also by the textured lines manufacturing. Different laser parameters combinations were used to assess the sintering and adhesion of HAp to the textured lines. HAp adhesion was assessed by performing high energy ultrasonic cavitation tests and sliding tests mimicking an implant insertion, with Ti6Al4V/HAp specimens sliding against animal bone. The HAp content retained after these tests was measured and results showed that an excellent HAp sintering and adhesion was achieved when using a scan speed of 1 mm/s and laser power between 9 and 9.6 W. It is important to emphasize that results indicated that the HAp bioactivity was maintained when using these conditions, validating this functionalization process for the production of hip prosthesis with improved bioactivity.por
dc.description.publicationversioninfo:eu-repo/semantics/publishedVersionpor
dc.description.sponsorshipFEDER -Federación Española de Enfermedades Raras(POCI-01–0145-FEDER-006941)por
dc.distributioninternationalpor
dc.identifier.doi10.1002/jbm.b.33964por
dc.identifier.issn1552-4973por
dc.identifier.pmid28748671por
dc.identifier.urihttps://hdl.handle.net/1822/52132
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherWileypor
dc.rightsrestrictedAccesspor
dc.subjectAdhesionpor
dc.subjectHydroxyapatitepor
dc.subjectLaser sinteringpor
dc.subjectLaser texturingpor
dc.subjectTi6Al4Vpor
dc.subject.fosEngenharia e Tecnologia::Engenharia Mecânicapor
dc.subject.wosScience & Technologypor
dc.titleTi6Al4V laser surface preparation and functionalization using hydroxyapatite for biomedical applicationspor
dc.typearticlepor
dspace.entity.typePublicationen
oaire.citationEndPage1545por
oaire.citationIssue4por
oaire.citationStartPage1534por
oaire.citationVolume106por
sdum.journalJournal of Biomedical Materials Research - Part B: Applied Biomaterialspor

Ficheiros

Pacote original

A mostrar 1 - 1 de 1
A carregar...
Nome:
41 - JBMRPA-AB-Far-Abr-Buc-Dou-Car-Sil-2017.pdf
Tamanho:
1.36 MB
Formato:
Adobe Portable Document Format