VHC performance evaluation at constant flow: 30 L/MIN

dc.contributor.authorOliveira, Ricardo Falcãopor
dc.contributor.authorMarques, Helena Maria Cabralpor
dc.contributor.authorMachado, A. V.por
dc.contributor.authorTeixeira, J. Carlospor
dc.contributor.authorTeixeira, S. F. C. F.por
dc.date.accessioned2018-03-23T11:19:24Z
dc.date.issued2016
dc.date.updated2018-03-22T09:48:18Z
dc.description.abstractAsthma treatment provided by a pressurized Metered Dose Inhaler (MDI) coupled to a Valved Holding Chamber (VHC), in cases of children younger than 5 years old, is a standard well-stablished in the medical community. The lack of experimental studies for comparison of several commercial VHC alternatives is the main goal of this study. The VHC device needs to be evaluated in terms of Fine Particle Mass (FPM) emitted and Performance. Such assessment was made based on a cascade impaction methodology (i.e. Multi-Stage Liquid Impinger - MSLI) at 30 ± 5 L/min. This impactor apparatus is composed by five stage of impaction, where the spray particle size distribution is collected by ranges of aerodynamic diameter, from coarse to fine particles. This evaluation was executed for 8 VHC devices: Aerochamber Plus®, A2A Spacer®, Compact Space Chamber Plus®, Space Chamber Plus®, Nebuchamber®, Vortex®, OptiChamber Diamond® and Volumatic®. The Ventolin® (salbutamol sulphate 100 μg/dose) was the chosen MDI device, due to its widely prescription and acceptance. Drug mass per stage was quantified by UV-Vis Spectrometry, through washing solutions of NaOH 0.01M. Results show clear distinction between the use of MDI alone or coupled with any VHC. Its Emitted Dose (ED) is higher when coupled to a VHC, although the FPM emitted is not different. In other hand the use of VHC provides a reduction from 82.7% to 95.3% of the Throat deposition in comparison to MDI alone. Results point the Aerochamber, Volumatic and Nebuchamber as the highest FPM emitters (27.5 ± 2.4 μg, 27.3 ± 2.7 μg, 26.5 ± 1.8 μg) and with high performance indexes (5.3 ± 0.1, 6.6 ± 1.3, 5.0 ± 0.4). A complete device characteristics analysis is provided, showing that Throat deposition is highly related to the valve design. The Leaflet design (≈5.0 μg) has lower throat deposition than Duck type (≈9.5 μg). A Handling and Attractiveness index is calculated and plotted against the Performance index divided by the device cost. This analysis shows that Volumatic is the best investment (1.73 ± 0.34 £−1) for home / hospital use (2.0 g−1·m−3), while the Aerochamber shows the best portability (53.2 g−1·m−3) and a reasonable investment (1.10 ± 0.02 £−1).por
dc.description.publicationversioninfo:eu-repo/semantics/publishedVersionpor
dc.description.sponsorshipThe first author would like to express his acknowledgments for the support given by the Portuguese Foundation for Science and Technology (FCT) through the PhD grant SFRH/BD/76458/2011. This work was financed by National Funds-Portuguese Foundation for Science and Technology, under Strategic Project UID/SEM/04077/2013 and UID/CEC/00319/2013.por
dc.distributioninternationalpor
dc.identifier.doi10.1115/IMECE201552283por
dc.identifier.isbn9780791857380por
dc.identifier.urihttps://hdl.handle.net/1822/53304
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherAmerican Society of Mechanical Engineers (ASME)por
dc.relationinfo:eu-repo/grantAgreement/FCT/SFRH/SFRH%2FBD%2F76458%2F2011/PTpor
dc.relationinfo:eu-repo/grantAgreement/FCT/5876/147456/PTpor
dc.relationinfo:eu-repo/grantAgreement/FCT/5876/147280/PTpor
dc.rightsclosedAccesspor
dc.subjectValved Holding Chamber (VHC)por
dc.subjectFine Particle Masspor
dc.subjectPerformance Indexpor
dc.subjectAsthmapor
dc.subjectExperimental evaluationpor
dc.subject.wosScience & Technology
dc.titleVHC performance evaluation at constant flow: 30 L/MINpor
dc.typeconferencePaperpor
dspace.entity.typePublicationen
oaire.citationConferenceDate13 - 19 Nov. 2015por
oaire.citationConferencePlaceHouston, Texaspor
sdum.bookTitlePROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2015, VOL 3
sdum.conferencePublicationProceedings of the ASME 2015 International Mechanical Engineering Congress and Exposition IMECE2015por
sdum.event.titleASME 2015 International Mechanical Engineering Congress and Exposition IMECE2015por
sdum.event.typecongresspor
sdum.export.identifier4693

Ficheiros

Pacote original

A mostrar 1 - 1 de 1
A carregar...
Nome:
10.1115@IMECE2015-52283.pdf
Tamanho:
364.22 KB
Formato:
Adobe Portable Document Format