Optical coupling between scintillators and standard CMOS detectors

dc.contributor.authorRocha, J. G.
dc.contributor.authorMinas, Graça
dc.contributor.authorLanceros-Méndez, S.
dc.date.accessioned2012-04-17T15:50:24Z
dc.date.available2012-04-17T15:50:24Z
dc.date.issued2006
dc.description.abstractIn the digital X-ray imaging systems, one of the main problems to be solved is the efficiency of the optical interfaces between scintillators and photodetectors. This article presents the theoretical analysis, simulations, and experimental results on two different optical interfaces (SiO2 and Si3N4), in order to maximize the X-ray detection efficiency. The working principle is the following: the X-ray photons reach the scintillator, which produces visible light. The visible light is then absorbed by the CMOS photodetector. Due to the fact that the refractive indexes of the scintillator and the photodetector are different, the visible light produced by the scintillator is partially reflected by the surface of the photodetector. In order to minimize this problem it is necessary to cover the photodetector with an anti-reflective filter. The anti-reflective filters were fabricated using two different dielectric layers available in a standard CMOS process, with no increase of the production time and costs of the devicespor
dc.identifier.doi10.1016/j.nima.2005.10.107por
dc.identifier.issn0168-9002
dc.identifier.urihttps://hdl.handle.net/1822/18702
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherElsevierpor
dc.relation.publisherversiondoi:10.1016/j.nima.2005.10.107por
dc.rightsrestrictedAccesspor
dc.subjectOptical interfacepor
dc.subjectThin filmpor
dc.subjectX-raypor
dc.subjectScintillatorpor
dc.subject.wosScience & Technologypor
dc.titleOptical coupling between scintillators and standard CMOS detectorspor
dc.typearticlepor
dspace.entity.typePublicationen
oaire.citationEndPage286por
oaire.citationIssue1por
oaire.citationStartPage281por
oaire.citationTitleNuclear Instruments and Methods in Physics Research Apor
oaire.citationVolume556por
sdum.journalNuclear Instruments and Methods in Physics Research Apor
sdum.publicationstatuspublishedpor

Ficheiros

Pacote original

A mostrar 1 - 1 de 1
A carregar...
Nome:
NuclearInstrumMPR A V556 (2006) pp281-286.pdf
Tamanho:
232.56 KB
Formato:
Adobe Portable Document Format
Descrição:
documento principal