Efficacy of a gas permeable contact lens to induce peripheral myopic defocus

dc.contributor.authorPauné, Jaumepor
dc.contributor.authorQueirós, A.por
dc.contributor.authorFerreira, Daniela Lopespor
dc.contributor.authorRibeiro, Miguel António Fariapor
dc.contributor.authorQuevedo, L.por
dc.contributor.authorGonzález-Méijome, José Manuelpor
dc.date.accessioned2016-01-18T17:11:39Z
dc.date.available2016-01-18T17:11:39Z
dc.date.issued2015
dc.date.updated2015-12-26T21:54:09Z
dc.description.abstractPurpose. The purpose of this work was to evaluate the potential of a novel custom-designed rigid gas permeable (RGP) contact lens to modify the relative peripheral refractive error in a sample of myopic patients. Methods. Fifty-two right eyes of 52 myopic patients (mean [TSD] age, 21 [T2] years) with spherical refractive errors ranging from j0.75 to j8.00 diopters (D) and refractive astigmatism of 1.00 D or less were fitted with a novel experimental RGP (ExpRGP) lens designed to create myopic defocus in the peripheral retina. A standard RGP (StdRGP) lens was used as a control in the same eye. The relative peripheral refractive error was measured without the lens and with each of two lenses (StdRGP and ExpRGP) using an open-field autorefractometer from 30 degrees nasal to 30 degrees temporal, in 5-degree steps. The effectiveness of the lens design was evaluated as the amount of relative peripheral refractive error difference induced by the ExpRGP compared with no lens and with StdRGP conditions at 30 degrees in the nasal and temporal (averaged) peripheral visual fields. Results. Experimental RGP lens induced a significant change in relative peripheral refractive error compared with the nolens condition (baseline), beyond the 10 degrees of eccentricity to the nasal and temporal side of the visual field (p G 0.05). The maximum effect was achieved at 30 degrees. Wearing the ExpRGP lens, 60% of the eyes had peripheral myopia exceeding j1.00 D, whereas none of the eyes presented with this feature at baseline. There was no significant correlation (r = 0.04; p = 0.756) between the degree of myopia induced at 30 degrees of eccentricity of the visual field with the ExpRGP lens and the baseline refractive error. Conclusions. Custom-designed RGP contact lenses can generate a significant degree of relative peripheral myopia in myopic patients regardless of their baselin spherical equivalent refractive error.por
dc.description.sponsorshipJaume Paune´ has proprietary and financial interests in the manufacturing and distribution of lenses evaluated in this study. The remaining authors declare that they do not have any proprietary or financial interest in any of the materials mentioned in this article. This work was partially funded by Fundação para Ciência e Tecnologia, Portugal, Projects: PTDC/SAU-BEB/ 098392/2008 and PTDC/SAU-BEB/098391/2008por
dc.distributioninternationalpor
dc.identifier.citationPauné, Jaume; Queiros, Antonio; Lopes-Ferreira, Daniela; Faria-Ribeiro, Miguel; Quevedo, Lluisa; Gonzalez-Meijome, Jose Manuel. Efficacy of a Gas Permeable Contact Lens to Induce Peripheral Myopic Defocus, Optometry and Vision Science, 92, 5, 596-603, 2015.por
dc.identifier.doi10.1097/OPX.0000000000000582por
dc.identifier.issn1040-5488por
dc.identifier.pmid25875679por
dc.identifier.urihttps://hdl.handle.net/1822/39500
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherAmerican Academy of Optometry (AAOPT)por
dc.relationinfo:eu-repo/grantAgreement/FCT/5876-PPCDTI/98392/PTpor
dc.relationinfo:eu-repo/grantAgreement/FCT/5876-PPCDTI/98391/PTpor
dc.rightsopenAccesspor
dc.subjectMyopia progressionpor
dc.subjectGas permeable contact lenspor
dc.subjectPeripheral refractionpor
dc.subject.wosScience & Technologypor
dc.titleEfficacy of a gas permeable contact lens to induce peripheral myopic defocuspor
dc.typearticle
dspace.entity.typePublicationen
oaire.citationEndPage603por
oaire.citationIssue5por
oaire.citationStartPage596por
oaire.citationTitleOptometry and Vision Sciencepor
oaire.citationVolume92por
sdum.journalOptometry and Vision Sciencepor
sdum.publicationstatuspublishedpor

Ficheiros

Pacote original

A mostrar 1 - 1 de 1
A carregar...
Nome:
5.5.110. PAUNE_15_OVS (Amiopik RGP).pdf
Tamanho:
434.85 KB
Formato:
Adobe Portable Document Format