Development and validation of a multi-modal customized device to stimulate in vitro cell culture systems

dc.contributor.authorOliveira, Sofia
dc.contributor.authorMonteiro, Francisca Andreia Azevedo
dc.contributor.authorCatarino, Susana Oliveira
dc.contributor.authorHinckel, Betina B.
dc.contributor.authorSotiropoulos, I.
dc.contributor.authorLeal, Ana Isabel Neto Cardoso
dc.contributor.authorSilva, Filipe Samuel
dc.contributor.authorCarvalho, Óscar Samuel Novais
dc.date.accessioned2026-01-28T09:24:50Z
dc.date.issued2025-11-14
dc.description.abstractThis work focuses on the development and validation of a multi-modal stimulation device for in vitro cell culture systems. The device was designed to stimulate cells or tissues placed on 12-well culture plates. It is connected to customized software that controls the parameters of photobiomodulation (PBM) and ultrasound stimulation (US) through light-emitting diodes and piezoelectric disks, respectively. A wide range of stimulation protocols can be explored by modulating central frequency or wavelength, power density, and duration. Four different cell lines were used to validate the safety and functionality of the device. Human osteoblasts, chondrocytes, periodontal ligament fibroblasts, and mouse-derived neuronal cells were cultured and stimulated daily with ultrasound (1.0 MHz, 100 mW/cm2, 5 min), light (810 nm, 7.5 mW/cm2, 5 min) and combined stimuli. After three days, metabolic activity and proliferation were assessed. Different cell types demonstrated distinct biological responses to the stimuli, as osteoblasts and chondrocytes showed increased metabolic activity after combined stimulation or PBM, while the metabolic activity of human fibroblasts or neuronal-like cells was unchanged after three days. This highlights the importance of a rigorous optimization of stimulation protocols according to the target tissue. The safety of the device and its sterilization conditions were demonstrated as there was no cell death or contamination during in vitro stimulation. This work features a feasible, safe, and effective multi-modal stimulation device that can provide a wide range of stimulation protocols to better understand their effect on cells or tissues.eng
dc.description.sponsorshipThis research was supported by the national funds of Portuguese Foundation for Science and Technology (FCT) under the scope of the UID/4436/2025 project; by LABBELS – Associate Laboratory in Biotechnology, Bioengineering and Microelectromechanical Systems, LA/P/0029/2020; by the projects Stimcart (doi: 10.54499/PTDC/EME-EME/4520/2021) and BrainStimMap (10.54499/PTDC/EME-EME/1681/2021); and by the PhD grants UI/BD/09375/2020 and UI/BD/150951/2021.
dc.distributioninternational
dc.identifier.articlenumber113306
dc.identifier.citationS Oliveira, F Monteiro, SO Catarino, BB Hinckel, I Sotiropoulos, A Leal, FS Silva, Ó Carvalho, Development and validation of a multi-modal customized device to stimulate in vitro cell culture systems, Journal of Photochemistry and Photobiology B: Biology, Volume 273, December 2025, 113306 (DOI: 10.1016/j.jphotobiol.2025.113306)
dc.identifier.doi10.1016/j.jphotobiol.2025.113306
dc.identifier.eissn1873-2682
dc.identifier.issn1011-1344
dc.identifier.pmid41240725por
dc.identifier.urihttps://hdl.handle.net/1822/99400
dc.language.isoeng
dc.peerreviewedyes
dc.publisherElsevier
dc.relationAssociate Laboratory on Biotechnology, Bioengineering and microELectromechanical Systems [LA/P/0029/2020]
dc.relationUID/4436/2025
dc.relationPTDC/EME-EME/4520/2021
dc.relationPTDC/EME-EME/1681/2021
dc.relationUI/BD/09375/2020
dc.relationUI/BD/150951/2021
dc.relation.hasversionhttps://www.sciencedirect.com/science/article/pii/S101113442500209X
dc.rightsopenAccess
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subjectCell culture
dc.subjectIn vitro studies
dc.subjectLight emitting diodes
dc.subjectPhotobiomodulation
dc.subjectPiezoelectric disk
dc.subjectAcoustic stimulation
dc.subjectUltrasound stimulationpor
dc.subject.fosEngenharia e Tecnologia::Engenharia Eletrotécnica, Eletrónica e Informática
dc.subject.odsSaúde de qualidade
dc.titleDevelopment and validation of a multi-modal customized device to stimulate in vitro cell culture systemseng
dc.typearticle
dspace.entity.typePublication
oaire.awardNumberLA/P/0029/2020
oaire.awardTitleAssociate Laboratory on Biotechnology, Bioengineering and microELectromechanical Systems [LA/P/0029/2020]
oaire.awardURIhttps://hdl.handle.net/1822/97179
oaire.citation.issueDecember 2025
oaire.citation.volume273
oaire.citationVolume273por
oaire.funderIdentifierhttp://doi.org/10.13039/501100001871
oaire.funderNameFundação para a Ciência e a Tecnologia, I.P.
oaire.fundingStreamConcurso para Atribuição do Estatuto e Financiamento de Laboratórios Associados (LA)
oaire.versionhttp://purl.org/coar/version/c_970fb48d4fbd8a85
relation.isProjectOfPublicatione6453fda-bc44-4538-917a-a94b535634c4
relation.isProjectOfPublication.latestForDiscoverye6453fda-bc44-4538-917a-a94b535634c4
sdum.journalJournal of Photochemistry and Photobiology B: Biology

Ficheiros

Pacote original

A mostrar 1 - 1 de 1
A carregar...
Nome:
1-s2.0-S101113442500209X-main.pdf
Tamanho:
3.05 MB
Formato:
Adobe Portable Document Format