Lab-on-a-chip for measuring uric acid in biological fluids

dc.contributor.authorMinas, Graça
dc.contributor.authorMartins, Júlio S.
dc.contributor.authorWilson, Cristina Pereira
dc.contributor.authorLima, Cristóvão F.
dc.contributor.authorWolffenbuttel, R. F.
dc.contributor.authorCorreia, J. H.
dc.date.accessioned2005-06-27T14:09:50Z
dc.date.available2005-06-27T14:09:50Z
dc.date.issued2002-09
dc.description.abstractThis paper describes a lab-on-a-chip for measuring uric acid in serum, plasma or urine. The lab-on-a-chip is composed by two dies: one is fabricated in polystyrene and contains the microchannels and the other is fabricated through a CMOS process and contains the photodetector and readout electronics. The uric acid concentration is measured by using a mixture of 14 µl of infinityTM uric acid reagent with 0.25 µl of sample. The achieved sensitivity is 0.33 mg/dl (±0.6% of human being urine values), with a 1 mm lightpath. Using an optical absorption method, a maximum peak at wavelength λ = 494 nm, is detected.eng
dc.description.sponsorshipFundação para a Ciência e Tecnologia - SFRH/BD/1281/2000, POCTI/33747/ESE/1999. FEDER.eng
dc.distributioninternationaleng
dc.identifier.citationRIPKA, P.; SANEISTR, J., ed. lit. – “Eurosensors XVI : proceedings of the european conference on solid-state transducers, 16, Prague, 2002”. [Prague?]: Czech Technical University, 2002. p. 66-69.eng
dc.identifier.urihttps://hdl.handle.net/1822/2353
dc.language.isoengeng
dc.peerreviewedyeseng
dc.rightsrestrictedAccesseng
dc.subjectLab-on-a-chipeng
dc.subjectOptical absorptioneng
dc.subjectUric acideng
dc.titleLab-on-a-chip for measuring uric acid in biological fluidseng